Skip to main content

Table 1 Limits of pressure imposed at the centre of the sample according to sample size for the most commonly used cavitron rotor

From: Measuring xylem hydraulic vulnerability for long-vessel species: an improved methodology with the flow centrifugation technique

Sample length (mm)

Centrifuge model

Maximum imposed pressure/rotation speed (Mpa/rpm)

Minimum imposed pressure/rotation speed (Mpa/rpm)

Maximum sample diameter (mm)

160

Sorvall R C series

− 0.2/2300

− 6/13,000

10

270

Sorvall RC series

− 0.2/1500

− 22/15,400

10

400

Beckman Coulter J6

− 0.5/1500

− 13/8100

21

1020

Custom

− 0.7/700

− 13/3000

24